Model : YK-80AP
FEATURES
* Microprocessor circuit assures maximum * The portable anemometer provides fast,
possible accuracy, provides special functions accurate readings, with digital readability and
and features. the convenience of a remote sensor separately.
* Large LCD with Dual function meter's display. * Multi-functions for air flow measurement :
* Heavy duty & compact housing case. m/s, km/h, ft/min, knots. mile/h.
* Records Maximum and Minimum reading * Thermistor sensor for Temp. measurement,
with RECALL. fast response time.
* Data hold. * Used the durable, long-lasting components,
* Auto shut off saves battery life. including a strong, light weight ABS-plastic
* Operates from 006P DC 9V battery. housing case.
* RS 232 PC serial interface. * Wide applications: use this anemometer to
* Low-friction ball vane wheels is accurate in check air conditioning & heating systems,
both high & low velocities. measure air velocities, wind speeds,
* Build in temperature ℃, ℉ measurement. temperature...etc.
GENERAL SPECIFICATIONS
Circuit Custom one-chip of micro- Over load Indicated by "- - - -".
processor LSI circuit. indication
Display * 51 mm x 32 mm supper large Operating Meter :
LCD display. Temperature 0 ℃ to 50 ℃(32 ℉ to 122 ℉).
* Dual function meter's display. Vane Probe :
Measurement m/s (meters per second), 0 ℃ to 60 ℃(32 ℉ to 140 ℉).
km/h (kilometers per hour), Operating Less than 80% RH.
ft/min (feet/per minute), Humidity
knots (nautical miles per hour), Power Supply 006P DC 9V battery
mile/h(miles per hour), (heavy duty type).
Temp.- ℃, ℉., Power Current Approx. DC 6 mA.
Data hold. Weight 220 g/0.48 LB.
Sensor Air velocity : Dimension Main Instrument :
Structure Conventional twisted van 207 x 68 x 29 mm
arm and low friction ball ( 8.1 x 2.7 x 1.1 inch).
bearing design. Sensor head :
Temperature Thermistor. Round, 72 mm Dia.
Memory Recall Records Maximum and Accessories Instruction manual........... 1 PC.
Minimum readings with RECALL. Included Anemometer probe.............1 PC.
Power Off Auto shut off saves battery life Carrying case, CA-06..........1 PC.
or manual off by push button. Optional RS232 cable, UPCB-02.
Sampling Time Approx. 0.8 sec. Accessories USB cable, USB-01.
Data Output RS 232 PC serial interface. Application windows software,
SW-U801-WIN.
ELECTRICAL SPECIFICATIONS (23 ± 5℃)
Measurement Range Resolution Accuracy
m/s 0.4 - 25.0 m/s 0.1 m/s ± ( 2% + 0.2 m/s )
km/h 1.4 - 90.0 km/h 0.1 km/h ± ( 2% + 0.2 km/h )
mile/h 0.9 - 55.9 mile/h 0.1 mile/h ± ( 2% + 0.2 mile/h )
knots 0.8 - 48.8 knots 0.1 knots ± ( 2% + 0.2 knots )
ft/min 80 - 4930 ft/min 1 ft/min ± ( 2% + 20 ft/min )
Temperature(℃) 0 ℃ to 50 ℃ 0.1 ℃ ± 0.8 ℃
Temperature(℉) 32 ℉ to 122 ℉ 0.1 ℉ ± 1.5 ℉
Note:
m/s - meters per second km/h - kilometers per hour ft/min - feet/per minute
mile/h - miles per hour knots - nautical miles per hour (international knot)