PRESSURE/VACUUM ADAPTER
Model : PS-403
SPECIFICATIONS
Range Pressure psi 0.5 to 500 psi
kPa 3.447 to 3447 kPa
Vacuum in-Hg 0 to 29.99 in-Hg
cm-Hg 0 to 76 cm-Hg
Accuracy kPa 3.5 to 1724 kPa :± (1 % + 3.5 kPa)
1725 to 2413 kPa : ± (2 % + 3.5 kPa)
* After Calibration 2414 to 3447 kPa : ± (5 % + 3.5 kPa)
* Temp. 23± 5 ℃ psi 0.5 to 250 psi : ± (1 % + 0.5 psi)
251 to 350 psi : ± (2 % + 0.5 psi)
351 to 500 psi : ± (5 % + 0.5 psi)
cm-Hg 0-76 cm-Hg : ± (5 % + 1.3 cm-Hg)
in-Hg 0 to 29.99 in-Hg :± (1 % + 0.5 in-Hg)
Output kPa 1 kPa/1 mV
Signal psi 1 psi/1 mV
cm-Hg 1 cm-Hg/1 mV
in-Hg 1 in-Hg/1 mV
Zero adjust. Adjust Potentiometer on the front panel.
Pressure 1/4" PS
Port Connector
Meter input 10 Meg ohm (If the meter input impedance is 1 Meg ohm, then ad 0.5%
Impedance to basic accuracy specification.)
Max. working Pressure 500 psi
Temperature Coefficient 1% Span (0 to 50 ℃)/max.
of Span
Operating Temperature Humidity
Temperature/ -10 to 10 ℃ 0 to 90% R.H.
Humidity 11 to 30 ℃
31 to 40 ℃ 0 to 75% R.H.
41 to 50 ℃ 0 to 45% R.H.
Measured The Pressue/Vacuum Adapter is compatible with industrial gases and
Materials liquids that are compatible with 316 stainless steel. To determine the
Compatibility compatibility of a liquid or gas, refer to manufacture's specifications.
When making measurements, be make sure that liquids or gases in those
systems are compatible. If they are not, contamination or damage to the
Pressure Adapter may result.
Battery Standard 9V battery (006P, 6F22 or NEDA ± 1604), heavy duty type.
Power Consumption Approx. DC 1.2 mA.
Weight 285 g/0.63 LB - included pressure sensor.
Size Main instrument : 10 x 5 x 2.5 cm
Pressure Sensor : 55 mm x 30 mm Dia.
Cable Length Approx. 1 meter/3.3 feet.
Accessories Operation manual............................................... 1 PC
Included 500 psi pressure transducer................................. 1 PC